|
|||||
|
|
|||||
| Home - Caratteristiche editoriali - Direttori e segreteria scientifica - Istruzioni per gli autori | |||||
|
|
|||||
Shuold we routinely stent small coronary arteries ?
Helmut Schühlen, MD, FESC; Adnan Kastrati, MD, FESC
| Introduction |
|
Introduction |
Coronary
stents are an frequently used for treatment of symptomatic coronary artery
disease. Their efficacy was shown in two pivotal trials, the BENESTENT1
and STRESS study2
. In both trials however, the protocol included only vessels of 3 mm
or larger diameter by visual estimate. Consequently, stenting was only
recommended as a treatment option for larger vessels3
. However, smaller vessels constitute a large group in daily practice of
percutaneous coronary interventions. Our own database indicates that 1/3 of
patients have lesions located in vessels <2.8 mm by quantitative measurements4
. Recently, several randomized trials comparing elective stent placement
with balloon angioplasty in small vessels have been completed. This review will
summarize these findings to address the impact of stent placement on long-term
outcome in small coronary vessels.
Why
are small coronary arteries different ?
Small
vessel size has been identified as an independent risk factor for the
development of restenosis after balloon angioplasty without stent placement5,6
. Several retrospective analyses have compared long-term results of stent
placement in small versus large vessels, indicating that restenosis rates differ
significantly with vessel size. Elezi et al. found that restenosis rate
increased from 20.4% in vessels >3.2 mm to 28.4% in vessels 2.8-3.2 mm,
and to 38.6% in vessels <2.8 mm (p<.001)4
. Similar findings were described by Akiyama et al., with a rate of 19.9%
in vessels ³3 mm,
and 32.6% in vessels <3 mm (p=.0001)7
. Both studies show that this negative effect is the result of simple
arithmetics: first, the initial acute lumen gain after stent placement in
smaller vessels is significantly lower than in large vessels. Second, both
trials have found that late lumen loss after six months is higher in small
vessels than in larger vessels. Both effects lead to an even more pronounced
decrease in net lumen gain, with a significantly higher loss index and
restenosis rate after six months. In large vessels, significantly lower
restenosis rates have been found after stent placement than after balloon
angioplasty in the BENESTENT and STRESS studies1,2
. In these trials, late lumen loss was also significantly higher after
stenting in comparison to balloon angioplasty. However, the final positive net
effect of stent placement was due to a significantly higher initial lumen gain
with a higher net lumen gain. In view of the above-mentioned findings in small
versus large vessels, data from trials in large vessels should not be
extrapolated to smaller vessels.
|
Results from Randomize Trials |
Within
the last few years, several dedicated randomized trials on small vessel stenting
have been performed. The findings from these studies are summarized in Table 1,
and discussed in more detail below.
The
first published randomized trial had been conducted in Korea by SW Park et al.8
. In this study, 120 patients with de-novo, non-ostial lesions with
reference diameter <3 mm by visual estimate were randomly assigned to
either balloon angioplasty of placement of 7-cell NIR stents (Boston-Scientific/Scimed).
Stent placement was successful in all 60 patients assigned to stent placement;
12 of 60 patients assigned to PTCA were crossed over to stent. Vessel size was
similar in both groups (2.48±0.24 mm
in PTCA group, 2.55±0.16 mm
in the stent group). The final minimal lumen diameter achieved was significantly
smaller in the PTCA group (2.14±0.37 mm)
than in the stent group (2.42±0.36 mm;
p<0.05). However, at follow-up angiography (obtained in 92.5% of eligible
patients), this initially observed benefit was reversed, with a minimal lumen
diameter of 1.53±0.48 mm
in the PTCA group, and 1.33±0.68 mm
in the stent group; restenosis rates were 30.9 and 35.7%, respectively. Due to
the low number of patients randomized, this difference failed to reach
statistical significance.
In
the ISAR-SMART (Intracoronary Stenting or Angioplasty for Restenosis
Reduction in Small Arteries) trial9
conducted at three centers in Germany, 404 patients with lesions in
native vessels between 2.0 and 2.8 mm by visual estimate were randomly
assigned to PTCA (n=200) or placement of a MultiLink stent (Guidant ACS).
Stenting was performed in 16.5% assigned to PTCA, and in 95.6% of patients
assigned to stent placement. Vessel size was 2.37±0.27 mm
(PTCA group) and 2.41±0.25 mm
(stent group; p=ns). In this study, the final minimal lumen diameter achieved
was also significantly smaller in the PTCA group (1.98±0.41 mm)
than in the stent group (2.35±0.40 mm;
p<0.001). At follow-up angiography (available in 83% of patients), the lumen
diameter was similar in the two groups (1.28±0.62
and 1.35±0.73 mm,
respectively; p=ns), as well as the restenosis rate (37.4 and 35.7%; p=ns).
The
BESMART (BeStent in Small Arteries) trial was a French multicenter study10
, utilizing the BeStent Small (Medtronic) for de-novo lesions in arteries
<3 mm by visual estimate. A total of 381 patients were randomized, 189
to PTCA, 192 to stent placement; 22.7% of patients in the PTCA group crossed
over to stents, in 1.0% of patients in the stent group no stent was placed.
Vessel size was 2.24±0.34 mm
in the PTCA group, and 2.23±0.36 mm
in the stent group (p=ns). Similar to all other studies, the acute angiographic
result was better after stent placement, with a minimal lumen diameter of 1.70±0.46 mm
after PTCA compared to 2.06±0.412 mm
after stent placement (p=0.001). However, this study yielded better results
for stent placement after six months, with a significantly lower minimal lumen
diameter after PTCA (1.19±0.58 mm
vs. 1.43±0.53 mm
in the stent group; p=0.0001), and higher restenosis rate (47 vs. 21%,
respectively; p=0.0001).
The
SISA study (Stenting in Small Arteries) was a multicenter worldwide study
performed in 352 patients with de-novo lesions in 2.3 to 2.9 mm vessels11-13
. The stent used for this trial was also the BeStent (Medtronic). The
cross-over rate was low in the stent arm (3.0%), but high in the PTCA arm
(20.9%). The reference diameters were 2.45±0.34 mm
(PTCA) and 2.50±0.37 mm
(stent; p=ns). In parallel to the other studies, the acute minimal lumen
diameter achieved was smaller after PTCA (1.82 vs. 2.28 mm in the stent
group). After six months, the minimal lumen diameter was similar in both groups
(1.38±0.56
and 1.44±0.53 mm,
respectively; p=ns), with no significant difference in restenosis rates (32.4
and 28.0%).
The
RAP study (Restenosis en Arterias Pequeñas) was performed in several
centers in Spain14
, comparing PTCA with stent implantation in de-novo lesions (vessel size
2.2 to 2.7 mm) in 426 patients. Like BESMART and SISA, this study was
performed with the small BeStent (Medtronic). Vessel size was 2.29±0.39 mm
(PTCA) and 2.32±0.32 mm
(stent; p=ns). The final lumen diameter was 1.84±0.62 mm
and 2.33±0.86 mm,
respectively (p<0.01). After six months, there was still a significant
difference in lumen diameter (1.31±0.58 mm
vs. 1.47±0.50 mm;
p=0.01), with a significant difference in restenosis rates (37 vs. 27%;
p<0.05).
Two
additional randomized trials have to be interpreted with caution as both were
stopped prematurely by the sponsors.
The
CORDIS-MICA trial was performed in several centers in the US, utilizing
the MiniCrown stent (Cordis)15
. With 600 patients initially mandated by the protocol, the trial was
stopped by the sponsor after randomization of 128 patients. The trial had a very
high cross-over rate of 37% in the PTCA arm. Like other studies, the acute
angiographic results were inferior in the PTCA arm (p<0.001). Based on a very
low number of repeat angiographies mainly driven by repeat symptoms, restenosis
rates were high but not significantly different (63 vs. 61%).
The
SISCA trial (Stenting in Small Coronary Arteries) was a multicenter trial
with the heparin-coated BeStent (Medtronic) performed in Norway and Sweden16
. The study was terminated by the sponsor after inclusion of 145 of
initially 200 planned patients. With a significant difference in the acute final
lumen diameter (1.79 vs. 2.22 mm; p<0.001), results after six months
were not significantly different (minimal lumen diameter, 1.57 vs. 1.69 mm;
restenosis rate, 19 vs. 10%).
How
should these conflicting results be interpreted ?
Detail angiographic data from all trials are summarized in Table 1 for direct comparison. Although a significant difference in the rate of angiographic restenosis had been observed in only two studies (BESMART and RAP), the restenosis rate in the PTCA arm was higher in all trials except for the study of Park et al.. To analyze this trend in favor of routine stenting further, we calculated the relative risk for angiographic restenosis using all data from all seven studies: this was highly significant with 0.74 (95% confidence interval, 0.66 - 0.83; p<0.001). However, several differences between these studies and specific aspects outlined below qualify the primary conclusion from this cumulative analysis.
Most
important, acute final results achieved were inferior in the PTCA group as compared
to the stent group in all studies. The smallest differences in final %-diameter
stenosis between the PTCA and the stent arm were observed in the Park study and
ISAR-SMART, the highest in SISA and SISCA. Suboptimal results after PTCA are an
independent predictor of restenosis5
. The highest discrepancies between acute angiographic results could be
observed in the PTCA groups, ranging from 13.7 to 29%. This difference between
studies might partially be responsible for differences in late angiographic
outcome. Figure 1 illustrates the absolute difference in restenosis rates
between the stent and PTCA arm of each study as a function of the final
%-diameter stenosis achieved in the PTCA arm. There is a clear linear
relationship (r=0.88), signifying that a positive effect of stenting was only
observed in the presence of poor acute results in the PTCA arm. In large vessels,
suboptimal results after PTCA are a valid indication for stent placement17
. Furthermore, studies on provisional stenting versus routine stenting in
large vessels have suggested that the benefit of routine stent placement
diminishes if PTCA yielded optimal angiographic results18-23
. Such a relationship is clearly present in small vessels, as illustrated
by Figure 1.Stent design is known to influence restenosis development24-26
, and these trials were performed with several different stent types. The
study by Park et al. with the 7-cell NIR stent, ISAR-SMART with ACS MultiLink,
BESMART, SISA, RAP and SISCA with the BeStent for small vessels, and CORDIS-MICA
with the MiniCrown stent. In a randomized comparison of five different stent
designs including the MultiLink, NIR and Palmaz-Schatz stents, we found a
significant difference in event-free survival after one year25
. In this unselected patient population (mean vessel size of 3.00 mm),
the lowest restenosis rate was observed for the MultiLink (25.3%), the highest
for the NIR stent (35.0%). These rates are lower than the respective rates of
the stent arms of ISAR-SMART (35.7%) and the study of Park et al. (35.7%) which
is consistent with known increasing risk for restenosis with decreasing vessel
size4,6
. The data from BESMART, SISA, RAP and SISCA indicate lower restenosis
rates for the BeStent in comparison to the other studies with other stent types.
There is only few data available from other trials for the BeStent. One small
observational study in an unselected study population found a restenosis rate of
27%27
. Moer et al. reported on a small randomized trial comparing the
BeStent to the 7-cell NIR stent in small coronary arteries28
. They found a restenosis rate of 22.5% with no significant differences
between the BeStent and NIR. This rate is comparable to the rates for the
BeStent from BESMART, SISA and RAP, but not with the rate for the NIR stent in
the study by Park et al. which was 35.7%. This indicates that the absolute
differences in restenosis rates between the different stent types in different
studies have to be interpreted with caution, and do not allow for a direct
comparison.
Most
of the trials used “dedicated” small vessel stents, (except ISAR-SMART with
the MultiLink). These stents have fewer struts, cells or loops per circumference
than large-vessel stents, and/or have thinner stent struts to reduce the metal
coverage and load relative to the circumferential vessel wall. This practical
approach however, has not been validated in dedicated clinical trials. In
contrary, there are data indicating that less coverage might lead to higher
restenosis rates: Garasic et al. have found that neointimal thickness is higher
for stents with eight struts per cross section as compared to stents with 12
struts per cross section in an experimental setting29
. Thinner struts on the other hand, have been associated with a lower
restenosis rate: in a randomized trial with 651 patients, a significantly lower
restenosis rate (15.0%) was found for a thin-strut stent (ACS MultiLink; strut
thickness 50 mm)
in comparison to a thick-strut stent (25.8%; p=.003), the MultiLink Duet (140 mm)30
; other than strut thickness, these two stents are almost identical in
design. Strut thickness has an influence on mechanical strength, and radial
force of different stents has been found to vary considerably31
. Furthermore, different mechanical characteristics may lead to
differences in elastic recoil after deployment32
. As geometric configuration of stents and vascular injury determine
restenosis development33
, further studies are needed to fully evaluate all modifications of
mechanical characteristics of stents on vessel trauma and restenosis development
in small vessels.
The
risk for restenosis increases with decreasing vessel size4,6
. Mean vessel size in the studies mentioned above was in the range of
2.23 to 2.69 mm. However, from these mean values there is no detectable
trend for increasing restenosis. For example, one of the lowest restenosis rates
were reported for the study with the smallest mean vessel size (BESMART).
Intravascular ultrasound (IVUS) imaging has provided additional information on
vessel size and dimensions. Frequently, the extent of coronary artery disease by
angiography and IVUS are discrepant34
with angiography underestimating the magnitude of disease. IVUS
measurements of vessel size and final lumen dimensions have been found to be
powerful predictors of angiographic restenosis35
. Furthermore, IVUS measurements have indicated that true vessel size
may be underestimated, ie, “small” vessels by angiography are frequently
diffusely diseased “large” vessels. Okabe at al. have further elucidated
this relationship36
: 176 patients with stent placement were divided into three groups
according to angiographic and IVUS measurements: I, large vessels by
angiography (³2.5 mm),
II, small by angiography (<2.5 mm), but large by IVUS (³4 mm),
and III, small by angiography and IVUS (<4.0 mm). Interestingly,
restenosis rates were not significantly different between group I and II (25.9
and 24.0%, respectively), but significantly higher in group III (66.7%, p<.05
in comparison to either I or II)36
. It can be hypothesized that vessel trauma was greater in group III than
in group II, leading to more neointimal hyperplasia. None of the randomized
trials in Table 1 utilized IVUS measurements; therefore their diverging data
cannot be explained by this hypothesis. Additional studies are needed to further
elucidate these findings.
Suboptimal
PTCA results are an established indication for stent placement in larger vessels.
The cross-over rate in the PTCA arm of the studies listed in Table 1 was in the
range of 14 to 23% (only exception CORDIS-MICA with 37%). These rates are higher
than in the earlier trials of BENESTENT (5.1%)1
and STRESS (6.9%)2
, but lower than in trials comparing strategies of optimal PTCA with
elective stenting (37 to 51%)22,23
. Therefore, the practice in these studies reflect a comparison of systematic
stenting versus PTCA with stents only for suboptimal results rather
than systematic stenting versus plain PTCA. The reasons for
cross-over from PTCA to stent placement in these studies are not available in
detail. However, typical reasons are suboptimal angiographic results, such as
residual lumen narrowing >50% and large dissections. Residual dissections are
a strong predictor of major cardiac complications during the early course37,38
, and to a lesser degree, final lumen diameter38
. Several multivariate analyses for the risk of early adverse cardiac
events or stent thrombosis during the first 30 days have not been able to
identify vessel size or reference diameter as an independent predictor37-39
; however, in most of these studies there is a trend for higher event
rates by univariate analysis or in direct comparisons of small with large
vessels4,7
. For studies comparing stenting with PTCA as listed in Table 1, clinical
event rates after 30 days are not available except for ISAR-SMART and SISCA. In
both trials, no statistically significant difference was found: in ISAR-SMART,
the event rate was 2.9% in the stent arm, and 1.5% in the PTCA arm, in SISCA,
the rates were 3.9 and 8.8%, respectively.
ISAR-SMART
is the only study with systematic use of glycoprotein IIb/IIIa inhibitors in
both arms. Results from various studies on the use glycoprotein IIb/IIIa
inhibitors during coronary interventions have yielded inconsistent results with
respect to the effect on angiographic restenosis40
, and there are no data indicating a contrasting effect on restenosis
after PTCA versus stent placement. Therefore, differences in the use of
glycoprotein IIb/IIIa inhibitors cannot account for the conflicting results of
the trials in Table 1 with respect to restenosis. However, it appears that the
low 30-day clinical event rate of ISAR-SMART in comparison with SISCA (see above)
could be well explained by the systematic use glycoprotein IIb/IIIa inhibitors
in ISAR-SMART.
![]() |
Table
1 - Angiographic Data from Randomized Trials
|
![]() |
Figura 1. This graph illustrates the correlation of the difference in restenosis rates (data for PTCA arm - data for stent arm) with the final %-diameter stenosis achieved in the PTCA arm at the end of the procedure. Each data point represents one of the study listed in Table 1: 1, Park et al.; 2, ISAR-SMART; 3, BESMART; 4, SISA; 5, RAP; 6, SISCA; 7, CORDIS-MICA. There is a clear linear correlation (r=0.88), suggesting that a benefit of systematic stenting in randomized small vessel trials was only present if procedures in the PTCA arm were finished with suboptimal final results. |
| Summary |
Available
data from randomized trials on stenting versus PTCA in small vessels do not
corroborate a general strategy of systematic stent placement in small vessels,
despite the fact that cumulative data suggest a significant benefit. However,
this advantage is clearly dependent on the final result achieved with plain PTCA.
Plain balloon angioplasty is clearly inferior if finished suboptimally, as
illustrated in Figure 1. It is notable that a strategy of plain PTCA required
stent placement in 10-20% of patients due to suboptimal angiographic results
after balloon angioplasty in all studies. Consequently, provisional stenting
after an aggressive attempt to achieve a good result with plain ballon
angioplasty seems warranted.
| References |
|
References |